User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Lab-Chron 1405:Timer DATE: 31-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 81 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 5335A STANDARD: Stanford Research FS700 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 HEAD * INITIAL CONDITIONS * 1.004 DISP The following additional equipment is required for 1.004 DISP this procedure: 1.004 DISP Timer Calibration Fixture 1.005 DISP Connect the UUT AC plug into the 1.005 DISP appropriate 115VAC receptacle. 1.006 DISP Press the UUT red power on button one time. 1.007 OPBR Is the UUT power on lamp (red pushbutton lamp) lit? 1.008 JMPT 1.011 1.009 DISP Press the UUT red power on button one time. 1.010 JMP 1.007 1.011 DISP Disconnect the UUT from the AC outlet. 1.012 DISP Connect the UUT AC plug to the Timer Calibration Fixture 1.012 DISP "TIMER" AC receptacle. 1.013 DISP Set the Fixture's OPEN/SHORT switch to the OPEN 1.013 DISP position. 1.014 DISP Plug Timer Calibration Fixture AC plug into the 1.014 DISP appropriate 115VAC receptacle. 1.015 DISP Press the UUT RESET pushbutton for a UUT reading 1.015 DISP of 00000. 1.016 DISP Connect the Timer Calibration Fixture as follows: 1.016 DISP TO COUNTER ........................... 5335A INPUT A 1.016 DISP VOLTAGE IN ........................... 5700A OUTPUT 1.017 STD Stanford Research FS700 1.018 HEAD {* Timer Calibration *} 1.019 JMP 2.005 1.020 EVAL 2.001 DISP Set the Fixture's OPEN/SHORT switch to the OPEN 2.001 DISP position. 2.002 DISP Press the UUT RESET pushbutton for a UUT reading 2.002 DISP of 00000. 2.003 IEEE [@5335]RE 2.004 JMP 2.009 2.005 IEEE [@5335][TERM LF] 2.006 IEEE [@5335]IN[D2000] 2.007 IEEE [@5335]FN2, AF1, CO1, TR1, AS1, AT-0.08, BT-1.37; 2.008 IEEE [@5335]WA1; 2.009 5700 5V S 2W 2.010 DISP Set the Fixture's OPEN/SHORT switch to the SHORT 2.010 DISP position. 2.011 OPBR Is the counter's GATE light on? 2.012 JMPF 2.001 2.013 HEAD TEST RUNNING - PLEASE WAIT 2.014 IEEE [D30000][D30000][D30000][D30000][D30000][D30000] 2.014 IEEE [D30000][D30000][D30000][D30000][D30000][D30000] 2.014 IEEE [D30000][D30000][D30000][D30000][D30000][D30000] 2.014 IEEE [D30000][D30000][D30000][D30000][D30000][D30000] 2.014 IEEE [D30000][D30000][D30000][D30000][D30000][D30000] 2.014 IEEE [D30000][D30000][D30000][D10000] 2.015 HEAD * Timer Calibration * 2.016 DISP Set the Fixture's OPEN/SHORT switch to the OPEN 2.016 DISP position. 2.017 ACC 1000.000T 0.001% 2.018 IEEE [@5335][I] 2.019 MEME 2.020 MEMI Use keyboard to enter actual UUT reading in Seconds: 2.021 MEME 2.022 MEMC T 1.0U #! Test Tol 1, Sys Tol 0.0001, TUR 10000.000 (>= 4.00). 3.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.